Capacitor and inductor analyzer
The Sencore LC103 takes the guesswork out of capacitor and inductor testing. It is a complete, automatic, microprocessor-controlled analyzer designed for service, maintenance or high-volume testing in the laboratory or at incoming inspection.
It identifies defective capacitors without removing them from the circuit. Using exclusive tests, it measures capacitor value and ESR, checks for parallel components that could skew the measurement, and gives a GOOD/BAD/SUGGESTED REMOVAL evaluation.
LC103 specifications
In-circuit tests (capacitors and inductors)
- Dynamic in-circuit tests to determine whether the component is good or bad.
- Component types: electrolytic, double-layer electrolytic, tantalum, ceramic and other capacitors.
- Yokes, flybacks, switching transformers and coils.
- Inductor range: 3.18 uH to 3.18 H
- Capacitor range: 0.002 uF to 20,000 uF
- Accuracy: same as the out-of-circuit tests for good inductors and capacitors with no parallel current paths.
- Suggested removal indication: triggered when the AC test differs from the DC test by more than 20%, or when the low-level DCR test indicates a leakage path greater than 20% of the charging current for capacitors.
Capacitors (out of circuit)
- Value: the dynamic capacitance test is determined by applying a constant current to the capacitor and measuring dV/dt.
- Range: 1 pF to 20 F
- Accuracy: ±1%, ±1 pF, ±1 digit for values up to 1990 uF, ±5%; 0.1% of full-scale range for values from 2000 uF to 20 F
Equivalent series resistance (ESR)
- Accuracy: ±5%, ±1 digit
- Capacitor range: 0.01 uF to 20 F
Dielectric absorption
- Accuracy: ±5% of reading, ±1 count
- Range: 1 to 100%
- Capacitor range: 0.01 uF to 20 F
Leakage
- Accuracy: ±5%, ±1 digit
- Applied voltage: keypad entry from 1.0 to 1000 volts in 0.1-volt steps; short-circuit current limited to 900 mA
Inductors (out of circuit)
- Value: dynamic test determined by measuring the emf produced when a changing current is applied to the coil under test.
- Range: 0.1 uH to 20 H
- Accuracy: ±2%, ±1 digit, ±0.1 uH
Ringing test
- Dynamic test of inductor quality: an excitation pulse is applied and the cycles the inductor rings before reaching a preset damping point are counted.
- Inductor range: 10 uH and above, non-iron core
- Accuracy: ±1 count on readings between 8 and 13 Rings
- Resolution: ±1 count
- Excitation pulse: 5 volts peak
General
- Operating temperature range: 32° to 104° F (0° to 40° C)
- Temperature range for specified accuracy (after 10 minutes of warm-up): 50° to 86° F (10° to 30° C)
- Power: 105-130 V AC, 60 Hz, 30 watts with the supplied PA251 power adapter.
- Operation on optional BY289 rechargeable battery.
- Automatic shutoff: cuts power in battery operation if the unit stays idle for more than 5 minutes.
