We have software and hardware to measure and characterize liquids and the effect of electric fields on them.
The Keysight 16452A is a fixture for measuring the dielectric permittivity and dielectric loss (tan δ) of liquids and low-viscosity materials. It works with impedance bridges or LCR analyzers, such as the Keysight E4980A, 4284A or similar.
Applications
- Characterization of liquid materials (dielectric oils, solvents, liquid polymers, etc.).
- Evaluation of liquids in research, quality control and R&D.
- Analysis of dielectric variations as a function of temperature or frequency.
Operating principle
The 16452A measures the impedance of the liquid held between two concentric electrodes in a test cell. That impedance is used to calculate:
- Relative permittivity (εr)
- Loss tangent (tan δ)
These parameters are derived from the applied frequency and the geometric constant of the cell.
The fixture uses the parallel plate method: the liquid sits between two electrodes and forms a capacitor. An LCR meter or impedance analyzer measures the resulting capacitance. It is designed to operate with 4-terminal-pair LCR meters or impedance analyzers.
Key technical features
| Parameter | Detail |
|---|---|
| Frequency range | 1 kHz to 1 MHz |
| Sample volume | Approximately 1.5 ml |
| Accuracy | High, determined by the connected bridge/analyzer |
| Electrode material | Stainless steel |
| Operating temperature | 15 °C to 35 °C (ambient) |
| Cell body | Chemically resistant plastic |
| Configuration | Vertical coaxial cell |
Other features
- Frequency: 20 Hz to 30 MHz
- Measurement parameters: capacitance (C) and dielectric constant (εr, εrr) of the liquid sample
- Sample quantity: ≤6.8 ml
- Weight: 6 lb (including the case)
Advantages of the 16452A
- High repeatability and accuracy in liquid measurements.
- Closed design that minimizes evaporation or contamination.
- Compatible with external temperature control systems for thermal testing.
- Easy to disassemble and clean.
Connectivity
It connects directly to the ports of an impedance bridge through low-loss coaxial cables. Keysight supplies the adapter cables for standard use.
Kit contents
- Coaxial test cell
- BNC or N-type connectors (depending on the model)
- User manual
- Assembly and cleaning tools