Gain and noise figure service by the Y method
We provide noise figure and gain measurements up to 18 GHz. We have noise sources ranging from 6 dB ENR to 21 dB ENR, to test a wide variety of devices.
We deliver graphs of the GAIN and NOISE FIGURE of your device. On request, we also deliver images or videos of the procedure.
What noise figure depends on
Noise figure depends mainly on these parameters:
- the frequencies used for the test;
- the receiver settings (for example, preamplifier, attenuator);
- the temperature during testing.
How to evaluate it correctly
- Measurements must be made across the receiver’s entire frequency range.
- The manufacturer must specify and publish a maximum noise figure value in the data sheet, for the receiver’s entire operating range. Since the values depend on frequency, the manufacturer can also specify the noise figure for selected frequency bands or ranges.
- A mean value (the arithmetic mean of a series of test measurements) can also be stated.
- Published values must be valid across the entire temperature range stated in the data sheet. Any limitations must be mentioned in the data sheet.
Basics of noise figure measurements
Noise figure is one of the main specifications of a monitoring receiver and is closely tied to its sensitivity.
The noise figure of a monitoring receiver is the factor by which the noise power delivered by the receiver increases when a reference noise is applied to the input. It is referred to the receiver input and measured at the output.
It can be measured by several methods:
- the “gain” method;
- the “Y-factor” method (noise source method);
- the “sensitivity” method.
Measurements must be made across the entire frequency range, tuning the receiver with test signals of frequencies f1, f2, … fn. Per octave, at least two frequencies are distributed evenly over the receiver’s entire frequency range.