Agilent 16191A SMD Test Fixture Rental

Test fixture for evaluating the impedance of side-electrode SMD components.

Agilent 16191A SMD test fixture

The Agilent 16191A test fixture is designed for impedance evaluation of side-electrode SMD components. The minimum SMD size it accommodates is 2(L)[mm].

Applicable instruments

4287A, 4297A, 4292A, 4395A with 4395A-010, 43961A, 4396B with 4396B-010, 43961A, E4991A.

Compensation and measurement

Open and short compensation, combined with electrical length compensation, are recommended before measuring. First enter the electrical length of the fixture in the electrical length compensation function of the measuring instrument.

  • Open compensation is performed by separating the high and low electrodes from each other. The gap must be equivalent to the width of the DUT.
  • Short compensation is performed with the shorting bar set from option 16191A-010/701.

After open and short compensation, together with electrical length compensation, the DUT is inserted into the test fixture.