16039A Binding Post Fixture rental, max. 250 MHz

Test fixture for evaluating the impedance of axial or radial lead devices, from DC to 250 MHz.

16039A Binding Post Fixture test fixture for rent

This test fixture is designed to evaluate the impedance of devices with axial or radial leads. Two binding posts spaced 7 mm apart ensure optimum contact with the component leads.

Applicable instruments

4287A, 4294A + 42942A, 4395A with Opt.4395A-010 + 43961A, 4396B with Opt.4396B-010 + 43961A, E4991A, (4195A + 41951A, 4286A, 4291A, 4396A)

When used with the 16085B: 4263B, 4268A, 4279A, 4284A, 4285A, 4288A, (4192A, 4194A, 4263A, 4278A)

Specifications

  • Frequency: DC to 250 MHz
  • Maximum voltage: ±40 V peak max. (AC+DC)
  • Operating temperature: 0 °C to 55 °C
  • Compensation and measurement: open and short

Compensation and measurement

Compensations are recommended in combination with electrical length compensation before measurement. First, the electrical length of the fixture must be entered in the electrical length compensation function of the measuring instrument.

Open compensation is performed with nothing connected to the binding posts. Short compensation is performed with the supplied shorting plate. After the open and short compensations, together with the electrical length compensation, the device under test is inserted into the fixture.