What it measures
A pyranometer measures total solar irradiance, the sum of direct, diffuse and reflected radiation, on a flat surface, in watts per square meter. It responds to a wide band of the solar spectrum. It is the reference instrument for determining the available solar resource.
How it works
Thermal models have a black sensor that absorbs radiation and heats up; the temperature difference relative to a reference generates an electrical signal, normally in millivolts, proportional to irradiance. One or two glass domes protect the sensor from wind and rain and limit the spectral band. Other models use a silicon photodiode, with a faster response but more limited spectral sensitivity.
How it is installed and used
- Measurement plane: to evaluate an array, the instrument goes in the same plane as the modules, with the same tilt and orientation. For the resource, it is placed horizontally.
- Leveling: a tilt error changes the reading of direct radiation.
- No shade or reflections: away from structures that cast shadows on it or reflect light onto it.
- Clean, dry dome: dust, dew and birds alter the reading.
- Connection: shielded cable and data acquisition compatible with the model’s output signal, using the sensitivity on the certificate to convert to W/m².
- Valid calibration with traceability, and periodic recalibration.
What an engineer sees in practice
A thermal pyranometer responds with a delay of seconds, and a photodiode one almost instantly. So when comparing with module power, which reacts immediately, these differences in response explain discrepancies on days with passing clouds. Silicon modules respond to a narrower spectrum than a thermal pyranometer.
How to interpret the reading
A reading is useful with its time, plane and instrument condition. Reference cells are used when the aim is to measure the irradiance actually usable by a module of the same technology. Electrical work on solar arrays must be carried out by qualified personnel in accordance with current regulations.