24/7 emergencies 814 777 6921812 230 6079WhatsApp 811 786 0768
Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

What is a functional test (bump test) of a detector?

A functional test (bump test) briefly exposes a gas detector to a known concentration to check that its sensors respond and its alarms activate, but it does not adjust the calibration.

What it is

A functional test, commonly known as a bump test, consists of exposing the detector to a gas of known concentration for a short time and verifying two things: that each sensor responds and that the alarms activate. It is a check of operation, not of accuracy.

How it differs from calibration

Calibration adjusts the instrument’s response so that the reading matches the value of the reference gas; in addition to gas, it requires an adjustment procedure. A functional test changes nothing in the instrument: it only confirms that it reacts within an acceptable margin. A useful analogy: calibrating is adjusting the scale, and the functional test is placing a known weight on it and seeing that it reads roughly what is expected.

Why it is done

A detector can switch on and show zero with a degraded, poisoned or blocked sensor. That makes it dangerous, because there is no indication that it is not working. The functional test is the way to find this out before relying on the instrument, which is why it is done before each use or workday, according to the applicable procedure and the manufacturer’s instructions. It is also done after an impact, a drop or exposure to high concentrations.

How it is done in general terms

  1. A test gas is used with the right concentration and mixture for the sensors, within its expiry date.
  2. It is connected with the instrument’s cap or adapter and a stable flow, without diluting it with air.
  3. It is confirmed that each sensor’s reading rises and that the visual, audible and vibrating alarms activate.
  4. The result is recorded.

How to interpret the result

If the instrument responds within tolerance, it is considered fit for use. If any sensor fails to respond, takes too long or falls out of tolerance, the instrument is not used to protect anyone: it is calibrated and the test is repeated, and if it keeps failing it is removed from service and the sensor is replaced.

What it does not replace

A functional test does not replace periodic calibration or continuous reading during the work. Criteria for frequency, tolerance and recording must be taken from the manufacturer’s instructions and the applicable safety procedure. Entry decisions are made by qualified personnel.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer