What poisoning means
In a gas sensor, “poisoning” is the loss of sensitivity caused by a substance that occupies or alters the site where the detection reaction takes place. It is not the same as normal aging or a loss of calibration, although all three can look the same on the display: the instrument responds less than it should.
The case of the catalytic sensor
A catalytic combustible gas sensor works by oxidizing the gas on a hot element whose temperature changes and, with it, its resistance. Certain compounds, among them those containing silicon, such as some silicones and sealants, and those containing sulfur or lead, deposit on the catalyst and make it inactive. This is the classic case: after exposure to silicone vapors, the sensor stays on and reads zero in the presence of a gas it used to detect.
The case of the electrochemical sensor
Electrochemical sensors use an electrolyte and electrodes that react with the target gas. They are affected by exposure to very high concentrations, by extreme dryness or humidity that alters the electrolyte, and by corrosive or cross-sensitive gases. They degrade gradually and eventually become depleted.
Why it is so dangerous
The damage does not usually trigger an alarm. A poisoned sensor keeps a stable, plausible reading. The only way to know it has lost sensitivity is to test it with a gas of known concentration and compare the response with what is expected. That is why the functional test before using the instrument is the main defense, more than any display indicator.
What an engineer sees in practice
- A functional test that does not reach the expected response or takes a long time to get there.
- A calibration adjustment that requires more and more correction to reach the value of the reference gas.
- An instrument that fails right after being used near aerosols, lubricants, sealants or cleaners.
How to reduce the risk
Avoid exposing the detector to vapors from silicone products and aggressive solvents, perform the functional test at each use, respect the manufacturer’s calibration intervals and remove from service any sensors that do not respond. A sensor that fails the test is replaced, not “compensated”. In spaces with a risk to life, entry decisions rest with qualified personnel.