Why it is the first measurement
Oxygen is measured first because it affects everything else. Catalytic combustible gas sensors need oxygen to work, and an oxygen-free atmosphere can give misleading combustible readings. In addition, a lack or an excess of oxygen is a hazard in itself.
How the measurement works
Portable detectors generally use an electrochemical oxygen cell. The gas enters by diffusion or through a pump, reacts in the cell and generates a current proportional to the partial pressure of oxygen. The cell is consumed over time, so it has a limited service life and is checked regularly.
Procedure
- Prepare the instrument. Switch on in clean air, let it stabilize, perform the functional test and set the oxygen zero to the normal air value if the procedure calls for it.
- Measure from outside. With a probe and hose, the sampling line is inserted through the opening before anyone enters.
- Wait. The sample takes time to travel along the hose, and the sensor takes time to respond. Measuring without waiting gives readings of the previous air.
- Sample at levels. Top, middle and bottom, and along the length of the space if it is large.
- Continue during the work with the personal detector.
What can affect the reading
- Pressure and altitude: the sensor responds to partial pressure, so a change in pressure affects the reading even if the proportion does not change. It is advisable to calibrate or zero under conditions similar to those of use.
- Rapid temperature changes and extreme humidity.
- Depleted sensor: readings that drift or do not stabilize.
- Leaking hose that draws in outside air and falsifies the measurement.
How to interpret
A normal reading corresponds to ambient air; the space is acceptable only if the reading is within the range set by the applicable procedure. A low reading indicates consumption or displacement and requires ventilating and looking for the cause. A high reading indicates enrichment, perhaps from an oxygen leak, and requires eliminating ignition sources. If in doubt, verify with another instrument. Measurement and entry must be carried out by trained people, in accordance with current regulations.
This guide is for information only. Work involving electrical hazards, heights or hazardous atmospheres must be carried out by qualified, authorized personnel in accordance with applicable regulations.