Two operations that get confused
The functional test (or bump test) checks that the instrument reacts: gas of a known concentration is applied and it is verified that the reading rises to within tolerance and the alarms activate. It does not change any settings. Calibration adjusts the sensor response so that the reading matches the reference gas. The first answers “is it fit for use today?”; the second corrects the drift.
Inspection before the test
- Physical condition: housing, filters, sensors free of dirt or water, no visible impact damage.
- Battery with enough charge for the shift.
- Pump and sampling hoses free of blockages or leaks, if the instrument has them.
- Date of last calibration and status of the maintenance program.
- In-date reference gas, suitable for the sensors and certified.
Test steps
- Switch on the detector in clean air, outside the space and away from contaminants, and wait for it to stabilize.
- Check the zero: with no gas, toxic and combustible readings must be zero and oxygen must correspond to normal air.
- Apply the reference gas with the specified flow and adapter.
- Observe the response: reading within the manufacturer’s tolerance, reasonable response time and activation of visual, audible and vibrating alarms.
- Remove the gas and confirm that readings return to the normal condition.
- Record the result, the date, the instrument and the person.
How to interpret the result
If the response is within tolerance and the alarms work, the instrument is fit for the shift. If not, it is calibrated and the test is repeated. If it still fails, it is taken out of service. A slow sensor, one that takes longer than normal to respond, is a warning of aging or contamination even if it reaches the final value.
When it is repeated
The frequency is set by the company program based on the manufacturer’s instructions and the risk. It is also repeated after a hard impact, after exposure to a high concentration or if there are doubts about the reading. A check skipped because “it worked yesterday” leaves the decision to enter without support. These tests and the entry must be carried out by trained people, in accordance with applicable regulations.
This guide is for information only. Work involving electrical hazards, heights or hazardous atmospheres must be carried out by qualified, authorized personnel in accordance with applicable regulations.