How is the atmosphere assessed before entry?

It is assessed with a calibrated gas detector, testing the atmosphere from outside at different levels of the space, measuring oxygen first, then flammable gases and then toxic gases, and monitoring continues for as long as anyone is inside.

Why it is assessed

A confined space can have an atmosphere that the senses cannot detect: lack of oxygen, excess oxygen, flammable gases or toxic gases. Many accidents happen because someone goes in to help another person without having assessed the air. Prior assessment, not intuition, determines whether entry is possible.

Measurement order

  1. Oxygen: first, because other sensors, such as catalytic combustible gas sensors, need oxygen to give a valid reading. Both deficiency and enrichment are dangerous; excess oxygen increases the risk of fire.
  2. Flammable gases and vapors: expressed as a percentage of the lower explosive limit.
  3. Toxic gases: those expected from the process or previous contents, such as hydrogen sulfide or carbon monoxide.

How sampling is done

  • From outside, with a sampling line or a pump, before fully opening or entering.
  • At different heights, because gases lighter than air collect at the top and heavier ones at the bottom.
  • Allowing time for the reading to stabilize, since long lines delay the response.
  • At several points if the space has compartments, branches or obstructions.

The instrument matters

The detector must be in good condition, with current calibration and a functional test with a known gas before each use. A depleted or poisoned sensor can show a normal reading in a dangerous atmosphere. In addition, each sensor has cross-sensitivities and conditions of use that must be understood.

Ventilation and continuous monitoring

Forced ventilation helps keep the atmosphere safe, but it does not replace measurement. Work such as welding, solvent cleaning or sludge removal generates gases during the task, so continuous monitoring is maintained and the conditions for leaving are defined. An attendant outside, with communication and rescue means in place, is advisable.

What is seen in practice

A normal oxygen reading and zero combustible gas guarantee nothing if measurement was done only near the entrance. It is common for the bottom of a tank, where sludge remains, to have a different atmosphere.

Common mistakes

  • Entering with the detector switched on, without having measured from outside first.
  • Measuring at only one level.
  • Skipping the functional test beforehand.
  • Measuring once and then not monitoring during the work.
  • Attempting a rescue without equipment or training.

Entry into confined spaces must be carried out by trained, authorized personnel, following the procedure and applicable regulations.

This guide is for information only. Work involving electrical hazards, heights or hazardous atmospheres must be carried out by qualified, authorized personnel in accordance with applicable regulations.