Measuring principle
Rotational speed is the frequency at which a full revolution repeats. An optical tachometer emits light and counts the pulses reflected back by a reflective mark; a contact tachometer uses a wheel or cone that turns by friction against the shaft; a stroboscope flashes at adjustable intervals until the mark appears to stand still.
Procedure with an optical tachometer
- Place a piece of reflective tape on the shaft or rotating element, on a clean area and well adhered.
- Aim the beam perpendicular to the mark, at the distance indicated by the instrument, without other shiny surfaces entering the field.
- Wait for the reading to stabilize before recording it.
- Repeat several times and compare.
What to consider
- Slippage: with contact types, if the wheel slips the reading comes out low.
- Transmission ratios: the measured speed is that of the part being targeted; on belts, pulleys and gearboxes it is not the motor speed.
- Multiple marks: more than one mark produces a reading that is a multiple of the true value.
- Intense ambient light: it can saturate the optical sensor.
How to interpret the result
In induction motors the actual speed is slightly lower than synchronous speed and the slip increases with load; the reading can therefore also be used to estimate qualitatively whether a motor is working under load. In vibration analysis, the exact speed is the reference for locating the synchronous component (once per revolution) and its orders in the spectrum.
Safety
Measuring near rotating parts exposes you to the risk of entanglement. Avoid loose clothing and keep your distance with the optical device whenever possible, since it does not require touching the shaft.