The model
Every body with a temperature above absolute zero emits electromagnetic radiation. A blackbody is an ideal emitter: it absorbs all the radiation reaching it and, at each temperature, emits the maximum radiation possible at all wavelengths. Its spectrum depends only on its temperature, not on the material.
Three fundamental relationships
- Planck’s law: gives the spectral distribution of radiation as a function of wavelength and temperature.
- Stefan-Boltzmann law: the total power emitted per unit area is proportional to the fourth power of the absolute temperature. That is why a small increase in temperature gives a notable increase in radiation.
- Wien’s displacement law: the wavelength of maximum emission is inversely proportional to the absolute temperature. Hotter objects emit their peak at shorter wavelengths.
From the blackbody to real objects
A real object emits less than a blackbody at the same temperature. The fraction is called emissivity, a number between zero and one. In addition, part of the radiation reaching the camera is not emitted by the object but reflected from its surroundings. The camera measures total radiation and, using the emissivity and reflected apparent temperature it is given, converts it into temperature. If those data are wrong, so is the result.
Why it matters in practice
- It explains why matte and oxidized surfaces measure better than polished ones.
- It justifies the use of reference blackbodies for calibration.
- It explains that the camera works in a specific band and not with the whole spectrum.
- It is the reason the relationship between temperature and signal is not linear.
A nuance
The blackbody is a model. The laws apply strictly to it, and they apply to real objects with emissivity corrections.