Why does a blurry thermal image ruin a diagnosis?

Because in an out-of-focus image the energy of a hot spot is spread over several neighboring pixels, which lowers the measured temperature and can lead to misclassifying the severity of a fault.

What happens to the energy

A focused camera concentrates the radiation from each point on the object onto one pixel. When focus fails, the radiation from that point is spread over a wider area of the detector. The central pixel receives less energy from the hot spot and more from its cooler surroundings. The result: the temperature peak flattens and widens.

Consequences for the diagnosis

  • The maximum temperature is underestimated, so a serious fault can look minor.
  • The shape of the pattern changes, and the distinction between a point and an extended area is lost.
  • Comparison with other images loses its meaning, because the focus conditions change.
  • A report may classify as “monitor” something that actually required action.

Other causes of a blurry image

  • Movement of the camera or target during capture.
  • A distance greater than the detector’s resolution allows for that object.
  • A dirty or damaged lens.
  • Measuring inside the minimum focus distance.
  • An object too small for the available pixels, in which case the image may be in focus and still average poorly.

Common errors

  • Trusting autofocus without checking the result on the viewer.
  • Focusing on the background rather than the component.
  • Refocusing after capture: on many cameras this cannot be corrected later in the software.
  • Taking the image with a wide-angle lens far from the target and expecting an accurate measurement.

Recommended practice

Focus on the component of interest, move closer or use a suitable lens until the target occupies several pixels, hold the camera steady, and check sharpness before saving. On energized equipment, the safe distance is defined by the applicable procedure and observed by qualified personnel.