What is calibrated at the factory
A radiometric camera associates each pixel’s signal with a temperature through a calibration performed against reference sources. That relationship depends on the properties of the detector, the electronics, and the optics. If any of them changes, the relationship shifts.
Causes of drift
- Aging of the detector and electronics: the responses of the elements change little by little.
- Thermal cycling: moving from cold to warm environments stresses the components.
- Impacts and vibration: these can displace the optics or damage the detector.
- Lens contamination: dust, grease, or coating damage attenuate the signal.
- Use outside the operating range: extreme ambient temperature or exposure to direct sun.
Cameras with uncooled detectors correct part of these variations with an internal shutter that refreshes the image, but that correction does not replace a calibration.
What the engineer sees
Without anything looking wrong in the image, the temperatures shift: a known reference measures slightly lower or higher than before, or two different cameras disagree when measuring the same object. The image remains useful for comparing patterns, but absolute temperatures are no longer reliable.
How to detect it
- Periodically measure a reference source, such as a blackbody, and record the deviation.
- Compare with another camera on the same target under the same conditions.
- Follow the manufacturer’s recommendations on the recalibration interval.
What to decide from the result
If the deviation is smaller than the uncertainty the application requires, you continue; if not, the camera is sent for calibration or adjustment. The interval depends on the use and on how critical the decisions are, not on a universal figure.