How it works
A resonant cavity stores electromagnetic energy in modes with characteristic frequencies. Its resonant frequency and its quality factor, which measures how much energy it loses per cycle, are determined with a network analyzer. When a sample is inserted in a zone where the electric field is strong, the frequency drops in proportion to the permittivity of the sample and the quality factor decreases in proportion to its losses.
Perturbation method
If the sample is small relative to the cavity, perturbation theory applies: the frequency shift gives the real part of the permittivity and the change in quality factor gives the imaginary part. The relationship depends on the volume of the sample relative to that of the cavity and on the field distribution at the placement point. With large samples, the small-perturbation assumption is no longer valid and full field models are required.
Advantages
- Excellent sensitivity for low-loss materials, because a high-quality-factor resonator detects tiny changes.
- Only a small amount of material is needed.
- No direct electrical contact with electrodes.
Limitations
- It measures at one frequency, or at a few frequencies of different modes, not across a broad band.
- The sample must have a precise shape and dimensions, such as a rod, sheet or capillary tube, and it is positioned carefully.
- In high-loss materials, the quality factor collapses and the resonance broadens so much that it hinders the measurement.
- Temperature changes the dimensions of the cavity, so it is controlled.
What the engineer sees
On the analyzer display, a resonance peak appears that shifts and broadens when the sample is inserted. Repeatability depends on always placing the sample in the same spot and on keeping the cavity thermally stable.
When to choose it
When the priority is the loss tangent of a low-loss dielectric and measuring at discrete frequencies is acceptable.