Operating principle
The probe ends in an aperture where the electric fringing fields penetrate the material under test. The impedance seen from the line depends on the complex permittivity of the material. A network analyzer measures the reflection coefficient, and a model of the aperture, which may be capacitive or full-wave, converts that figure into permittivity and losses.
Materials that measure well
- Liquids: the probe is immersed, with perfect contact and no gaps.
- Soft solids and semisolids: pastes, gels, tissues.
- Rigid solids with a flat, polished face.
The sample must be thick enough that the field does not reach its rear face; otherwise, the measurement includes whatever is behind it.
Calibration
This is the decisive step. It is done at the aperture plane with three known references: open circuit (air), short circuit and a liquid of well-known permittivity, such as deionized water. The references are chosen so that their permittivity is close to that of the sample. Any cable movement after calibration alters the phase and degrades accuracy, which is why low-flex cables are used and handling them is avoided.
Sources of error
- Air gaps between the probe and a solid, which cause the permittivity to be underestimated.
- Surface roughness.
- A sample that is too thin.
- Very low-loss materials, whose effect on the reflection is tiny and falls within the noise.
- Temperature changes between calibration and measurement.
What the engineer sees
The probe offers a fast broadband measurement with little preparation. In exchange, its uncertainty is greater than that of resonant or filled-line methods, and repeatability depends on the pressure and the quality of the contact.
When it is not the best option
For very small loss tangents, or when high accuracy is needed on a solid, a resonant or transmission line method is preferable.