What we want to know
Permittivity changes with temperature because of the expansion of the material, the mobility of the dipoles and phase or glass transitions. Knowing this dependence shows how far, for example, the frequency of a resonator or an antenna will shift when the equipment heats up or cools down.
Typical setup
The sample, with its electrodes or inside its holder, is placed in a chamber whose temperature is controlled and measured close to the sample, not only in the chamber air. Cables and accessories passing through the hot zone must withstand the range and not introduce drifts of their own. For high-frequency measurement, the assembly is calibrated taking into account the changes in the fixture with temperature.
Measurement protocol
- Condition the sample, above all by drying it, before heating.
- Bring the temperature to a set value and wait for the sample to stabilize, since it has thermal inertia.
- Measure and move on to the next point; repeat on the way up and on the way down.
Phenomena that appear
- Hysteresis: if heating and cooling differ, there is slow relaxation, moisture loss or changes in the material.
- Loss peaks: these indicate dipolar relaxations or a glass transition.
- Expansion: alters the thickness and the contact with the electrodes, which must be corrected for or limited with bonded electrodes.
- Conduction: increases at high temperatures, and at low frequency it dominates the apparent losses.
How to express the result
It can be given as a curve of permittivity against temperature at a fixed frequency, or as a coefficient of variation over an interval where the behavior is almost linear. State the interval and the frequency, because the coefficient is not constant over the whole range.
What to check first
If the results look erratic, check the location of the temperature sensor, the stabilization time and the moisture state of the sample before suspecting the material.