Definition
In a thermal cycle, the sample is brought to a high temperature, held there, brought to a low temperature and held again. Each holding interval is the dwell. The key is the instant at which counting starts: it must begin when the sample, and not just the chamber air, reaches the target temperature within tolerance.
Why the sample lags behind
The temperature of a part follows that of the air with a delay that grows with its thermal mass. If time is measured from the air reading, the sample may spend little time at the extreme, or none. That is why thermocouples are placed at the slowest points, normally the most massive component, and the dwell is counted from the moment that point stabilizes.
What the dwell accomplishes
- Thermal stabilization: the entire part reaches the extreme and gradients disappear.
- Stress relaxation and creep: in materials such as solders, time at extreme temperature allows them to deform and accumulate damage. A short dwell can make the test less damaging per cycle.
- Function check: operation is sometimes measured during the dwell, when the temperature is stable.
Effect on the result
For the same range and rate of change, increasing the dwell usually makes the cycle more severe for time-dependent mechanisms such as creep, but has little effect on those governed solely by temperature amplitude. That is why the number of cycles to failure can change with it, and results with different dwells cannot be compared without care.
How it is defined
It is taken from the product specification or the test standard and verified against the actual response of the sample. The report states whether it was counted from the air or from the sample. A cycle with an excessively long dwell lengthens the test without adding information.
What to check first
Before starting, a verification cycle is run with the real load and the thermocouple curves are reviewed: Did the part reach the extremes? How long was it within tolerance? That evidence is what backs up the result.