What it is for
Solar radiation produces two distinct effects. One is thermal: surfaces absorb energy and heat up above the air temperature. The other is photochemical: ultraviolet radiation breaks bonds in polymers, paints and adhesives, causing discoloration, brittleness and loss of gloss. The chamber makes it possible to study both under repeatable conditions.
How the radiation is produced
Xenon arc lamps are used, which produce a spectrum close to that of the sun, or fluorescent UV lamps, which concentrate on an ultraviolet band. Optical filters adjust the spectrum to resemble direct sunlight or sunlight passing through glass. Irradiance is measured with a sensor and regulated to maintain the desired level.
What is controlled
- Total irradiance and irradiance in specific bands, particularly the ultraviolet.
- Air temperature and the temperature of a reference surface.
- Relative humidity and, in certain tests, condensation or water spray cycles.
- Uniformity of the irradiance over the samples.
An important distinction
An accelerated weathering test chamber studies the degradation of materials with cycles of light, humidity and heat. A solar radiation test on equipment mainly evaluates the heating and thermal effects of the sun on the product. These are different approaches, even though they share the lamp.
How to interpret the result
The correlation between hours of exposure and years outdoors is not universal. It depends on the material, the location and the season. Samples are therefore compared against a reference material exposed at the same time. Lamps age and lose intensity, so they are calibrated and replaced according to a schedule.
How it compares with a real environment
A chamber does not reproduce the full sun: it lacks daily variation, real rain and the contaminants of the site. It is therefore advisable to define in advance what you want to learn, whether the maximum heating of a dark surface or the rate of discoloration of a finish, and to choose the spectrum and temperature accordingly.
Precautions
The ultraviolet radiation from these lamps harms the eyes and skin; the chamber must be operated with the doors closed and with the protective measures provided by the manufacturer.