How temperature uniformity is measured inside a chamber

An array of calibrated sensors is placed in the working volume, the chamber is stabilized at the setpoint, and the difference between the hottest and coldest sensors at the same instant is calculated over a recording period.

What uniformity measures

Uniformity is the temperature difference between different points of the working volume at the same instant. It should not be confused with stability, which describes how much a point varies over time, or with accuracy, which compares the actual temperature with the setpoint. A chamber can be stable and accurate at the control sensor and still have zones that differ by several degrees.

How it is measured

  1. Define the working volume and the test temperatures.
  2. Choose the points: the corners, the center and the positions near the control sensor, with greater density in large chambers.
  3. Use calibrated sensors of the same type, with similar response and known uncertainty.
  4. Stabilize until no sensor shows a trend any longer.
  5. Record for a time that covers several control cycles.
  6. Calculate the difference between the maximum and minimum values at each instant, and the average of each sensor against the setpoint.

What the engineer sees

The data show zones that are systematically colder or hotter, for example near the door, the air return or the heaters. If the differences change with temperature, the cause is usually heat loss through the walls or air circulation. If they change with the load, the cause is blocked airflow.

Factors to control

  • The chamber load: empty or with a representative mass.
  • The fan speed.
  • Door opening during recording.
  • Thermal contact of the sensors with the air and not with surfaces.

What to do with the result

Uniformity is reported together with the measurement uncertainty and is used to decide where to place critical samples. The acceptance criterion depends on the standard or the test requirement. After repairing or moving the chamber, it is advisable to repeat the measurement.