The two ends of the response
The zero reading is the indication when the input is at the lower limit of the range; on a gauge pressure instrument, this is with the pressure equal to atmospheric. The full-scale reading is the indication when the input is at the upper limit. If both are correct and the response in between is linear, the instrument is properly adjusted.
What errors they produce
- Zero error: shifts the entire response equally. It is the easiest error to detect because it can be seen without applying a signal.
- Full-scale or span error: changes the slope. The error grows in proportion to the input and is greatest at the upper end.
- Non-linearity: the response departs from the straight line between the two points, with greater error at intermediate points even when the ends are correct.
How it is verified
Apply the zero input and record the output. Apply the full-scale input and record it again. Then check intermediate points, rising and falling, to detect linearity and hysteresis problems. Checking only zero and full scale is not enough, because an instrument can be correct at the ends and fail in the middle.
How it is adjusted
Zero is adjusted first and then full scale, and the cycle is repeated because on many instruments the two adjustments slightly affect each other. The zero adjustment is made with a truly nil, stable input; adjusting it with residual pressure or at a different temperature shifts the error to the rest of the scale.
What an engineer sees
On a pressure gauge, a pointer that does not come to rest at zero signals zero error or mechanical damage. On a transmitter with a standard current signal, zero and full scale correspond to the ends of the signal, and comparison against a calibrator shows how much of the error is offset and how much is slope.