What the triple point of water means

The triple point of water is the unique condition of temperature and pressure at which liquid water, ice and vapor coexist in equilibrium. Its temperature is a fixed reference point in thermometry.

Definition

The triple point of water is the state in which solid, liquid and gaseous water coexist in equilibrium. It occurs at a single combination of temperature and pressure; any other combination eliminates at least one of the phases. On the ITS-90 scale it is assigned the value of 273.16 K, and it is traditionally related to 0.01 °C.

Why it is so useful

It is highly reproducible: it depends on a pure substance, the equilibrium is easy to recognize, and its value does not depend on ambient pressure, unlike the freezing point in air, which does vary with pressure and with dissolved air. That is why calibrations of standard platinum resistance thermometers are referred to this point.

How it is realized

A sealed glass cell containing high-purity water and vapor, with no air, is used. A layer of ice is formed around a central well, and the cell is surrounded with a mixture of ice and water to maintain the temperature. The thermometer is inserted into the well and allowed to reach equilibrium. The cell is kept in a bath or insulated container to preserve the ice for hours.

What the engineer sees

In a laboratory, the reading in the cell is used to check a thermometer’s resistance at that temperature and to detect drift, before or after other calibrations. A standard thermometer that deviates at the triple point indicates that it has suffered damage or contamination.

Sources of uncertainty

  • Impurities in the water or the cell.
  • Isotopic composition of the water, which slightly affects the equilibrium temperature.
  • Hydrostatic pressure effect due to immersion depth.
  • Self-heating of the thermometer.
  • Insufficient thermal contact in the well.

What it is not

It should not be confused with an ice bath or with the melting point of water at atmospheric pressure. Those are useful references, but they have lower reproducibility than the triple point cell.