What happens inside the gauge
The Bourdon tube flexes with every change in pressure and moves the pointer through a toothed sector and a pinion. If the pressure oscillates rapidly, those parts work on every cycle: the joints and teeth wear, play develops, and the pointer movement becomes erratic or gets stuck because of deformed parts or dirt.
Where pulsation comes from
- Reciprocating or plunger pumps.
- Compressors.
- Rapid valve opening and closing, which produces water hammer.
- Unstable regulators or oscillating control loops.
How it shows up
The pointer vibrates so much that it cannot be read, the zero shifts because the mechanism has deformed, or the pointer stays fixed at one value even though the pressure changes. Over time, repeatability is lost and hysteresis increases. The window may fog if the seal fatigues.
What is done about it
- Damper or restrictor: limits the flow to the element and smooths out the peaks. If the restriction is excessive, it clogs with particles; if it is insufficient, it does no good.
- Liquid-filled case: glycerin or another fluid damps the vibration of the mechanism and lubricates the parts. It must be compatible with the service temperature.
- Diaphragm seal: isolates the element from aggressive or dirty fluid.
- Suitable range: keep the maximum pressure, peaks included, away from the top end of the scale.
- Attenuating the source: an accumulator or pulsation dampener in the line treats the cause.
A useful criterion
If the pointer visibly shakes, that is a sign that the gauge is living in an environment that will shorten its life. After strong pulsation events it should be verified against a standard, because a gauge that appears to work may have lost accuracy.