Why a thermocouple loses accuracy over time

A thermocouple drifts because its wire changes: the composition and structure of the alloys are altered by temperature, atmosphere and thermal cycling, and the voltage it generates no longer follows the table.

The root cause: the signal originates in the wire

In a thermocouple, the voltage is not generated only at the junction, but along the wire wherever a temperature gradient exists. If the wire becomes inhomogeneous, the voltage depends on where the gradients are, not just on the temperature at the tip. That is why moving the thermocouple within the process can change its reading.

What degrades it

  • Selective oxidation: at high temperature, one of the alloy components oxidizes more than another and alters the local composition.
  • Contamination: vapors, metals or reducing gases from the process diffuse into the wires and poison them.
  • Thermal cycling and mechanical stress: these alter the crystal structure and cause deformation that changes the response.
  • Ordering changes in the alloy, which depend on the exposure time within a certain temperature range.
  • Insulation degradation, which creates shunts and unwanted additional junctions.

What the engineer sees

Typical signs are a reading that slowly moves away from that of another sensor at the same point, a process that needs more energy to reach the same reading, or differences between redundant thermocouples. An unstable reading that jumps when the cable is moved points more to a damaged junction or compromised insulation than to drift.

How it is detected and managed

Periodic calibration compares the thermocouple with a standard at one or more points. A test at a single temperature does not reveal inhomogeneity that appears only with an immersion profile; for that reason, where it is critical, verification is done at different depths. Thermocouples exposed to severe service are replaced on a criterion based on drift history, not just on the calendar.

Practical conclusion

Drift cannot be corrected with an adjustment at the readout, because the error changes with the position of the gradient. The recommended approach is to select the type and sheath suited to the atmosphere, avoid exposing the thermocouple to the edge of its temperature limit, and replace it when the comparison calls for it.