General scope
ASME B40.100 is an American standard on indicating pressure gauges and their accessories that use an elastic element (Bourdon tube, diaphragm or bellows). It establishes a common language between manufacturer and user on how a pressure gauge is classified, built and specified.
Topics it organizes
- Accuracy grades: a letter-based classification indicating the permitted error as a percentage of span, differentiated by zones of the scale.
- Construction: case, element and connection materials, case types and overpressure protection.
- Dial: graduations, lettering and range, so that normal use falls in the middle part of the scale.
- Marking and accessories: such as seals, dampers and relief devices.
Specific values and classes change from one edition to the next, so the current document must be consulted before specifying.
Why it matters to the person operating the gauge
The standard explains why a high-grade gauge costs more and why operating near the ends of the scale is not recommended. It also underpins practices such as choosing a range that covers the maximum pressure with ample margin, using seal elements or diaphragms with corrosive or viscous fluids, and fitting damping where there are pulsations.
What an engineer sees
A specification sheet shows the accuracy grade, the element material, the dial size and the case type. When a gauge is calibrated, the acceptance criterion usually compares the deviation against the percentage of the stated grade, which is why it is tested at several points with rising and falling pressure to detect hysteresis.
When it is not enough
The standard describes the instrument, not its installation or the safety of the system. Selecting a compatible fluid, protecting against spikes and handling hazardous gases or fluids depend on the process and the applicable code. For formal decisions, the current edition must be consulted.