What it is in one sentence
ITS-90 is a practical scale designed to reproduce thermodynamic temperature as closely as currently possible, using reproducible physical states called fixed points and interpolation equations for the instruments between them.
Why it is needed
Thermodynamic temperature is not measured directly with an ordinary thermometer. The scale provides an agreed procedure so that laboratories in any country obtain the same values when realizing the same temperature. Without a common scale, two calibrations would not be comparable and traceability would be meaningless.
Its components
- Fixed points: phase transitions of pure substances, such as the triple point of water, the freezing points of metals and the triple points of other gases, each assigned a temperature value.
- Interpolating instruments: the standard platinum resistance thermometer is used over much of the range; other defined instruments are used elsewhere, such as gas thermometers at low temperatures and radiation thermometers at very high temperatures.
- Deviation functions: relationships that describe how the thermometer departs from a reference function and allow interpolation between fixed points.
How it relates to everyday measurement
An engineer does not realize the scale directly. It reaches the engineer through the traceability chain: a primary laboratory realizes it and calibrates standard thermometers, and these in turn calibrate probes and working baths. Each link adds uncertainty. When a certificate states traceability, it means its values can be linked to that scale through documented comparisons.
What to keep in mind
ITS-90 replaced earlier scales, so older data may differ slightly from current data in certain ranges. In addition, the scale is a practical approximation and not exact thermodynamic temperature; the difference is small, but acknowledging it is part of an honest uncertainty statement.