What is a dry-block calibrator?

A dry-block calibrator is a temperature-controlled metal block with holes, heated or cooled, into which sensors are inserted to calibrate or verify them without the need for liquids; it is practical and portable.

What it is

A dry-block calibrator, or dry-well calibrator, is a device that holds a metal block at a controlled temperature. The block has one or more wells into which the sensors are inserted. Because metal is a good conductor, it distributes the heat and keeps all the wells at a similar temperature.

How it is used

  1. Choose the insert with holes matching the diameter of the sensors, so that the fit is close.
  2. Set the temperature and wait for the block to stabilize.
  3. Insert the sensor and a reference thermometer and wait for both to reach the same temperature.
  4. Compare the readings.

Advantages

  • It uses no liquids: no spills, vapors or fluid maintenance.
  • Portable, and therefore suitable for on-site calibration.
  • Quick to heat and easy to use.

What limits quality

  • Axial uniformity. The temperature varies along the well, especially near the opening, because of losses to the surroundings. That is why immersion depth matters.
  • Differences between wells. The wells are not at exactly the same temperature.
  • Clearance between sensor and well. The air between them is a poor conductor. A sensor with a smaller diameter than the well makes worse contact and gives a larger error.
  • Load effect. A large sensor, or several sensors at once, alter the temperature of the block.
  • Stabilization time. Longer than that of a bath in some cases, especially with bulky insertions.

An example

A short sensor, with its element near the tip, does not reach the most uniform zone of the well if immersion is insufficient. The reading will be lower than that of a longer standard, and that difference can be mistaken for a sensor error.

When it does not apply

For irregularly shaped or very large sensors, or when the lowest possible uncertainty is required, a fluid bath or a specialized medium is usually a better option. The same is true when the sensor does not fit any insert.

Safety

The surfaces of the block and the sensors reach temperatures that pose a burn risk. It must be used by trained personnel, following the manufacturer’s instructions.