The principle
In a metal, as temperature increases, the vibrations of the crystal lattice intensify and the electrons meet more obstacles, so the resistance rises. Platinum is preferred because this relationship is very stable, repeatable and well characterized, and because the material is chemically inert. An RTD (resistance temperature detector) is an element of this kind, with a nominal value defined at zero degrees Celsius; a Pt100 has one hundred ohms at that temperature.
How it is read
The instrument passes a small current through it and measures the resulting voltage, calculates the resistance and converts it to temperature using a standardized curve or the sensor’s calibration coefficients. The relationship is not perfectly linear, so an equation is applied rather than a simple proportion.
Wiring: two, three or four wires
The sensor has low resistance, so the resistance of the leads matters.
- Two wires: the lead resistance adds to that of the sensor and appears as a temperature error.
- Three wires: compensates for the leads if all three are equal, using an additional measurement.
- Four wires: separates the current wires from the voltage wires and eliminates the effect of the wiring resistance.
Self-heating
The excitation current dissipates heat in the element itself. In still air, where dissipation is poor, the sensor can read higher than the actual temperature. That is why a low current is used and the medium in which the measurement is made is taken into account.
Compared with a thermocouple
An RTD is usually more stable and accurate in the moderate range, while a thermocouple covers higher ranges, responds faster and is more rugged. They are not interchangeable in every application.
What a technician observes
- Readings that stray from a reference because of poorly connected wiring or connection resistance.
- Slow drift after exposure to shock, vibration or extreme temperatures.
- Erratic values caused by moisture entering the sheath.
Common errors
- Using two wires with long leads and not compensating.
- Insufficient insertion of the sheath into the medium, with conduction error along the tube.
- Calibrating without stabilization time or without a uniform bath.
- Replacing the sensor without checking that the instrument uses the same curve.