How is a pressure decay leak test performed?

The test pressurizes the system, isolates it and monitors how much the pressure falls over a defined time, carefully distinguishing a drop caused by a real leak from one caused by temperature changes.

Principle

If a closed volume is pressurized and isolated, and there are no leaks, the pressure should hold. If it falls, something is escaping. The test is based on observing that drop over a period of time and comparing it with an acceptance criterion set by the applicable specification.

General sequence

  1. Prepare the system. Isolate the test volume, install a transducer or gauge of suitable resolution and verify that the system can withstand the test pressure.
  2. Pressurize. Raise the pressure to the test value in a controlled manner.
  3. Allow it to stabilize. Wait for thermal and mechanical effects to dissipate before starting the timing.
  4. Isolate the source. Close the supply valve.
  5. Observe. Record pressure and temperature for the set time.
  6. Evaluate. Compare the drop, corrected if applicable, with the acceptance criterion.

The great enemy: temperature

With gases, pressure depends on temperature. When a gas is compressed, it heats up; when it is isolated and cools toward ambient temperature, its pressure falls even if there is no leak. A drop due to cooling can be misread as a leak. Likewise, a temperature rise from the sun or from nearby equipment can hide a real leak. That is why you wait for stabilization and record the temperature, in order to correct for it or to decide whether the test is valid.

Gas and liquid

  • With gas, because of its compressibility, the drop is slower and depends on the volume: a large volume takes longer to reflect a small leak. A leak rate can be estimated from the pressure drop, the volume and the time.
  • With liquid, which is nearly incompressible, a small loss of volume produces a large pressure drop. For this reason, trapped air complicates the reading, since it behaves like a spring.

What the engineer sees

A drop that levels off over time is usually thermal. A continuous, sustained drop points to a leak. A drop that begins when the system is moved suggests a seal or connection that is sensitive to stress.

How to interpret the result

The test only demonstrates that the system complies within the sensitivity of the method: it does not detect leaks smaller than what the resolution and the time allow you to see. If the leak needs to be located, other methods are used, such as detection with soap solution or tracer gas.

Safety

A pressurized vessel stores energy; with gas, the stored energy is much greater than with liquid. The test must be performed by qualified personnel, following the applicable procedures and regulations.