The principle
A deadweight tester, or pressure balance, balances a pressure against a known force. A piston of known effective area slides inside a cylinder, and calibrated masses are placed on it. When the piston floats freely, the pressure of the fluid beneath it equals the force exerted by the masses divided by the effective area. It does not depend on an elastic element or on electronics, which is why it is a primary or very high-level pressure standard.
What is corrected
To obtain the actual pressure from the masses, you must consider:
- Local gravity, which varies with latitude and altitude.
- Air buoyancy on the masses.
- The temperature of the piston-cylinder assembly, because of thermal expansion of the effective area.
- Distortion of the effective area with pressure.
- The height difference between the piston reference level and the gauge, which creates a fluid column.
General procedure
- Level the tester and verify that the fluid is suitable for the gauge and the equipment.
- Connect the gauge, bleed the air from the circuit and check for leaks.
- Place the masses for the first point and generate pressure until the piston floats.
- Gently rotate the piston to reduce static friction and wait for it to stabilize.
- Record the gauge reading, tapping the dial lightly if the procedure calls for it.
- Repeat at increasing points, and then at decreasing points, to reveal hysteresis.
- Calculate the error at each point using the corrected pressure.
Points to watch
- The gauge height must be matched to the reference level, or corrected for.
- Trapped air must be avoided, because it dampens the system and gives a false impression of stability.
- The working fluid, oil or gas, must be compatible; a gauge used with oil must not later be exposed to oxygen.
How to interpret the result
The error at each point, together with the uncertainty of the tester and the corrections, is compared with the tolerance of the gauge. The ascending and descending points give an idea of hysteresis and friction.
Safety
High pressures store energy, and a connection failure can be dangerous. The work must be done by qualified personnel, with the applicable accessories and safety procedure.