How does the resistance of a metal vary with temperature?

The electrical resistance of a metal increases with temperature almost linearly over a moderate range, because the greater agitation of the crystal lattice scatters the conduction electrons more.

Direct answer

In a metal, resistance rises as temperature goes up. It is said to have a positive temperature coefficient. At higher temperature, the lattice atoms vibrate with greater amplitude, the conduction electrons collide with them more, and their net motion is hindered, so resistivity increases.

What the curve looks like

For small variations, the relationship approximates a straight line: resistance changes in proportion to temperature, with a coefficient characteristic of the material. Over wide ranges a curvature appears, which is described with a polynomial. For platinum sensors, standardized equations of the Callendar-Van Dusen type are used, which reproduce that curve with a few coefficients.

What contributes to resistance

  • Thermal part. Due to lattice vibrations; this is the part that changes with temperature.
  • Residual part. Due to impurities and defects; it is almost independent of temperature and dominates at very low temperatures.

That is why the purity of the metal matters: a wire with impurities has lower relative sensitivity to temperature and less reproducibility.

Why platinum is used

Platinum is chemically stable, resists oxidation over a wide range, can be obtained at high purity, and its resistance-temperature relationship is reproducible and well characterized. Other metals, such as nickel or copper, are used in specific applications, but with narrower ranges or lower stability.

How the resistance thermometer exploits it

A resistance thermometer (RTD) uses this property: its resistance is measured and, with the sensor equation, converted to temperature. Two practical considerations:

  • An excitation current must be passed to measure the resistance, and that current heats the element; this is self-heating.
  • The resistance of the leads adds to that of the sensor, so three- or four-wire connections are used to compensate for it.

Comparison with other materials

Semiconductors, such as thermistors, behave differently: in the most common ones, resistance falls as temperature rises and the change is much larger, though more nonlinear.

What the engineer sees

A reading that rises when the leads are touched or when the cable length is changed usually gives away uncompensated lead resistance. A sensor that drifts over time may have suffered contamination or mechanical stress in the element.