Definition
Tolerance is the difference between the upper and lower permissible limits of a quantity. It is set in the design or specification, according to what the product or process needs in order to work. It is an engineering decision, not a property of the measuring instrument.
What it is not
- It is not uncertainty. Uncertainty describes how well a measurement result is known. Tolerance describes how much the product is allowed to vary.
- It is not accuracy. The accuracy of an instrument is how close its readings are to the true value. Tolerance is a requirement.
- It is not the instrument’s maximum error. Although the maximum permissible error of an instrument is sometimes called “tolerance”, correct usage distinguishes between the two.
How it relates to measuring
To verify that a part is within its tolerance you have to measure it, and every measurement has uncertainty. Near a limit, uncertainty makes conformity doubtful: a part measured right at the limit may be inside or outside. That is why a decision rule is established: the part is accepted only if the result plus its uncertainty falls within the limit, or else it is accepted with a certain assumed risk.
Ratio to uncertainty
For the measurement to be useful, the uncertainty must be clearly smaller than the tolerance. When the two are of the same order, many of the parts fall in the zone of doubt and the cost of wrong rejections or acceptances rises. The measurement system is usually required to use only a fraction of the tolerance.
Ways of expressing it
| Form | Meaning |
|---|---|
| Symmetric bilateral | Equal deviation on both sides of the nominal |
| Asymmetric bilateral | Different limits on each side |
| Unilateral | One limit only, for example a maximum |
What the engineer sees
A tolerance too tight for the process produces rejects; one too wide produces a product that does not work. Reviewing the tolerance together with the capability of the process and of the measurement system avoids adjusting the wrong thing.