What Is a Reference Material?

A reference material is a material sufficiently homogeneous and stable with respect to one or more properties, with assigned values and their uncertainty, used to calibrate, validate methods or control the quality of measurements.

Some quantities cannot be embodied in a simple physical object, such as chemical composition, the hardness of a metal or the pH of a solution. For these, the reference is a material whose property is known: it is measured with your own method, and the result must agree with the assigned value within the uncertainties. This confirms that the method and the instrument are working.

What it must meet

  • Homogeneity: every portion of the material has the same property, within known limits.
  • Stability: the property is preserved over time and under the stated storage conditions.
  • Assigned value with uncertainty: determined by well-characterized methods, often by several laboratories.
  • Traceability: the value is related to a higher-order reference.
  • Documentation: a certificate describing the material, the value, the uncertainty and the instructions for use.

Reference material and certified reference material

A certified reference material comes with a certificate issued through a technically valid procedure that states the value, the uncertainty and the traceability. A non-certified reference material can be useful for internal control, but its value does not have the same backing.

What it is used for

  1. Calibration: establishing the relationship between the instrument’s response and the value of the property.
  2. Method validation: demonstrating that the method delivers results close to the known value.
  3. Quality control: measuring it routinely to detect deviations.

An example

In pH measurement, buffer solutions with an assigned value are used to adjust the meter; in metal analysis, alloys of certified composition are used to check a spectrometer; in hardness testing, reference blocks are used to verify the hardness tester.

When it does not apply

The material must resemble what is actually being measured, in matrix and in the level of the property. An expired or poorly stored material no longer guarantees its value.