It is a portable or benchtop unit that replaces several separate sources and meters. It is used to apply a known signal to an instrument and compare its response, or to measure the signal the instrument delivers. Its value depends on its accuracy and on being calibrated with traceability, because it becomes the working standard for the other instruments.
Typical functions
- Voltage and current source: for testing transmitters, indicators, recorders and controllers. The loop current signal is very common in industry.
- Measurement of the same signals: for reading the output of a transmitter.
- Resistance simulation: for testing resistance temperature detector inputs.
- Thermocouple simulation: generates the voltage corresponding to a temperature, with cold junction compensation.
- Frequency and pulses: for flow or speed sensors.
- Loop power: some supply voltage to power a transmitter during measurement.
What the engineer sees in practice
To calibrate a temperature transmitter, the sensor is disconnected, the calibrator is connected in its place to simulate the sensor, and the transmitter output is checked against the simulated temperature. This verifies the electronics, not the sensor or its installation. That is why this type of test does not replace a calibration of the sensor with a temperature source.
How to choose one
The accuracy of the calibrator must be clearly better than that of the instrument under test, with the corresponding uncertainty. Points to weigh are the signal types it covers, its ranges, resolution, stability and ruggedness for field use. It is advisable to confirm on the certificate which functions and points are calibrated, because they are not always all calibrated.
Limitations
A calibrator is only as good as its last calibration and its conditions of use. Cables, connections and temperature can add errors, especially with small signals such as those from thermocouples.