Temperature, pressure, humidity, vibration and electromagnetic interference can affect both the standard and the instrument. There are three ways of dealing with them, which are usually combined: control the environment, correct the result and declare the uncorrected part as uncertainty.
1. Control the environment
The most robust approach is to avoid the effect. Work is done in a space with stable temperature, equipment is allowed to reach thermal equilibrium with the environment before measuring, it is isolated from vibration and drafts, and it is protected from electric fields. Stabilization time is part of the procedure, not a luxury.
2. Correct the result
When the effect cannot be eliminated but its relationship with the quantity is known, a correction is applied. Examples:
- In length, the thermal expansion of the part and the standard is corrected using their expansion coefficients and the difference from the reference temperature.
- In pressure, the height between reference planes or the density of the fluid may require corrections.
- In mass measurement, air buoyancy has an influence that depends on its density and that of the masses.
- In electronic instruments, ambient temperature can affect drift, and the manufacturer’s coefficient is applied.
3. Include it in the uncertainty
Every correction starts from data that have uncertainty: the temperature is measured with a thermometer that has its own, and the coefficients are known within limits. In addition, the residual effect after correction is treated as a contribution. If it is not corrected, its maximum reasonable magnitude is estimated and added as a source in the budget.
How to interpret the result
A certificate must state the environmental conditions during calibration. If your use takes place under very different conditions, the results may not be representative, and it is advisable to apply your own corrections or to calibrate on site. Documenting the environment also makes it possible to explain differences between laboratories.