What is the probe attenuation factor?

The probe attenuation factor is the ratio between the voltage at the tip and the voltage reaching the oscilloscope input; the instrument uses it to scale the reading and display the actual voltage at the tip, which is why it must match the probe in use.

What it indicates

A ten-to-one attenuating probe delivers one tenth of the tip voltage to the oscilloscope. A unity-ratio probe delivers it unchanged. The factor expresses that ratio. There are also high-voltage probes with larger factors. For current probes, the equivalent is a sensitivity in volts per ampere.

Why the instrument needs it

The oscilloscope channel only knows the voltage reaching its connector. To display the tip voltage, the instrument multiplies the reading by the configured attenuation factor and adjusts the vertical scale and automatic measurements accordingly. If the configured factor does not match that of the probe, all amplitude readings are wrong by exactly that ratio.

How it is set

  • Automatic recognition: many probes have a contact or an identification circuit that tells the channel the correct factor.
  • Manual setting: the factor is selected in the channel menu. This is necessary with probes that have no identification.
  • Switchable probes: if the probe has a selector between attenuating and unity, the channel setting must also be changed every time the selector is moved.

An example of error

A signal is measured with an attenuating probe, but the channel is set to a unity factor. The oscilloscope shows an amplitude ten times smaller than the real one. The signal looks small and noisy, and amplitude and power measurements come out wrong by the same factor. Times and frequencies, however, are not affected.

What to check first

  1. The physical selector on the probe, if it has one.
  2. The factor shown in the channel menu.
  3. That compensation was adjusted with that same factor.
  4. That the maximum voltage of the probe is sufficient, because the factor does not increase it by itself.

With a known signal, such as the instrument’s calibration signal, you can verify that the displayed amplitude matches the expected one before measuring.