What is the rise time of an instrument?

The rise time of an instrument is the time its output takes to go from a low reference level to a high one when it receives an ideal step, and it is related to its bandwidth: the greater the bandwidth, the shorter the rise time.

Definition

If a perfect step is applied to the input of an instrument, the output does not rise all at once: it rises over a finite time that depends on the instrument’s response. Rise time is usually measured between two percentages of the transition amplitude, often ten and ninety percent, although other percentages are also used and you should know which ones were applied.

Relationship to bandwidth

For a first-order low-pass response, the product of the ten-to-ninety percent rise time and the minus three decibel bandwidth is approximately 0.35. This is a useful rule, not an exact equality: other responses, such as Gaussian responses or those of instruments with steeper filters, give somewhat different constants. The idea to remember is the inverse relationship: more bandwidth, faster edges.

What it means when measuring

What appears on the screen is the signal as modified by the measurement system. When you measure a signal whose rise time is comparable to that of the instrument, the measured value is greater than the real one. The rise times of cascaded elements, such as the signal, the probe and the oscilloscope channel, combine approximately as the root of the sum of the squares. This is why an instrument with a rise time much shorter than that of the signal alters the measurement very little, while one close to it distorts the measurement significantly.

In practice

  • If the measured edge has a rise time almost equal to that of the instrument, you are measuring the instrument’s response rather than the signal.
  • A long probe or ground connection increases the rise time of the system.
  • To evaluate an edge with good accuracy, the measurement system must be clearly faster than the edge.

When it does not apply

The constant rule does not describe well an instrument with a response very different from first order, nor does it work for signals without a well-defined edge. In those cases, consult the manufacturer’s characterization.