The mechanism
The ground lead supplied with a passive probe has inductance, which grows with its length and with the area of the loop it forms with the signal conductor. This inductance sits in series with the probe’s input capacitance. A fast edge contains energy at the resonant frequency of this combination and excites it, so a damped oscillation appears right after the transition. The longer the lead, the lower the resonant frequency and the closer it falls to the spectral content of the signal.
What you observe
- Overshoot and ringing after each fast edge.
- A rise time that differs from the real one and depends on the length of the ground lead.
- Changes in the waveform when you move or coil the lead.
- Signals picked up in the loop, such as switching from a nearby circuit.
On slow signals it is not noticeable, because the resonant frequency lies far above the signal content.
How to check it
- Measure the signal with the long ground lead.
- Repeat with a minimal ground connection, such as the spring that fits on the probe body.
- Compare. If the overshoot and ringing decrease greatly, they came from the loop.
How to correct it
- Remove the tip cover and use the ground spring or a short-tip adapter.
- Connect the return to a ground point close to the measurement point, not to a distant point on the board.
- For very fast signals, use an active probe with high-frequency tip accessories, or coaxial test points.
- Do not use the long lead with fast edges, convenient though it is, because the measurement is no longer reliable.
An example
A digital clock appears with strong overshoot and sustained ringing. With the short ground, the waveform becomes clean and the overshoot is small. The first image showed more of the probe than of the circuit. A design conclusion based on it would have led to adding unnecessary damping.