What problem it solves
The normal oscilloscope input measures a signal with respect to its own ground, which is also tied to protective ground. If you connect the probe to a point whose return is at another voltage, a current flows through the ground connection, the circuit can be altered and the instrument can be damaged. A differential probe amplifies only the difference between its two terminals and delivers to the oscilloscope a signal referenced to its ground, so the circuit under test is left free of that connection.
Differential mode and common mode
If one input is at V1 and the other at V2, the useful signal is V1 minus V2 (differential mode), and what they share is the common mode. An ideal probe ignores the common mode completely. In practice, part of the common mode appears as error at the output. The common-mode rejection ratio (CMRR) expresses how well it is attenuated, and it worsens as frequency rises, so a low-frequency specification does not guarantee good rejection on a fast edge.
What to look for in the specifications
- Maximum differential voltage: the limit between the two tips.
- Maximum common-mode voltage: each tip with respect to ground. This is a different limit from the previous one and the one most easily exceeded.
- Selectable attenuation: to match the range to the signal and preserve resolution.
- Bandwidth and CMRR as a function of frequency.
- Measurement category and voltage rating.
Where it is used
To measure across a power bridge, on floating gates in an inverter, on a current sense resistor away from ground, or on any network that has no node in common with the instrument. For a ground-referenced signal, on the other hand, a passive probe is usually simpler.
What to expect in practice
With a large common mode and fast edges, a spurious reading synchronized with the switching appears even if the differential signal is small. This is checked by joining both tips to the same node: what you then see is the common-mode error of the measurement. Measurements on energized circuits or circuits with hazardous voltages must be made by qualified people, following the applicable safety regulations and using accessories with the appropriate category and voltage rating.