How do you measure overshoot and settling?

Overshoot is how far the peak exceeds the final value of a transition, expressed as a percentage of the step amplitude, and settling is the time the signal takes to remain within a tolerance band around that final value.

Definitions

For a step that goes from an initial level to a final level, the amplitude is the difference between the two. Overshoot is the excess of the maximum peak over the final level, and its percentage is that excess divided by the amplitude. Settling measures how long the response takes, from the start of the transition, to enter and stay within a band around the final level. You define the band, so the time is meaningless unless its size is stated.

Procedure

  1. Capture a transition with a time base long enough to see the complete settling, including a stable portion after the event.
  2. Determine the initial and final levels as steady states of the signal, not as the highest or lowest peak. Instruments with automatic measurement use a histogram to estimate them.
  3. Measure the maximum peak relative to the final level with vertical cursors and calculate the percentage.
  4. Set the tolerance band and use horizontal cursors to locate the instant at which the signal enters it for the last time.
  5. Measure the time from the start of the step to that instant.

Settings that matter

  • Vertical scale: magnify the area near the final level using offset to see the band in detail. An instrument with low vertical resolution limits the narrowest band you can resolve.
  • Bandwidth: if the instrument or the probe is slow relative to the edge, the measured overshoot decreases and the edge lengthens.
  • Noise: averaging helps as long as the trigger is stable.

Measurement artifacts

A long ground connection on the probe adds oscillations of its own that are easily mistaken for circuit overshoot. Before attributing the overshoot to the design, check that it persists with a minimal ground connection and with another probe. The IEEE 181 standard defines these parameters in a standardized way, and consulting it helps when measurements must be compared between laboratories.

How to interpret the result

High overshoot with slowly decaying oscillation indicates low damping. Long settling with no overshoot indicates an overdamped system or one with a large time constant.