What floating means and why it is a problem
On a typical oscilloscope, the common of all the inputs is the same and is connected to ground. This is convenient for ground-referenced signals, but it prevents measuring between two points neither of which is at ground, as at the output of a transistor bridge, in a circuit powered from the mains without isolation, or in a switching power supply. Connecting the probe common to one of those points forces it to ground and produces a short-circuit current.
Correct alternatives
- Differential probe: it measures the difference between its two inputs, which need not be referenced to ground, and delivers it to the oscilloscope referenced to ground. It is selected by maximum differential voltage, maximum common-mode voltage, bandwidth and common-mode rejection.
- Oscilloscope with isolated channels: each channel has its own reference, galvanically separated from the others and from ground, which allows direct measurement at several different floating points.
- Two passive probes with math subtraction: this consists of measuring each point with respect to ground and subtracting. It is acceptable only if both points are at a voltage that the probes and the instrument can withstand with respect to ground, and its accuracy is poor because any difference between channels appears as error.
What must not be done
Removing the oscilloscope’s ground connection, or powering it through an adapter that omits it, so that it floats, is dangerous. The enclosure and input connectors can end up at mains voltage or another high voltage, within reach of anyone. Nor is it advisable to rely on an isolation transformer without understanding the risk, because the instrument can still have capacitance to ground.
What you see in practice
- With a differential probe, the signal is clean only if common-mode rejection is sufficient at the frequency of the variations; if not, an attenuated copy of the common-mode voltage appears on the signal.
- With isolated channels, the capacitance to ground of the isolation and long leads can couple high-frequency noise.
An example
To observe the gate voltage of a high-side transistor in a half bridge, which swings with respect to ground, it must be measured with respect to its own source: a differential probe or an isolated channel with good common-mode rejection is used.
These measurements involve hazardous voltages; they must be performed by qualified personnel in accordance with the applicable regulations.