Why it is needed
An attenuating probe forms a divider with the oscilloscope input, which is a resistance in parallel with a capacitance. For the attenuation to be the same at low and high frequency, the resistive divider and the capacitive divider must have the same ratio. The probe’s adjustable capacitor matches that ratio. Because input capacitance varies from one channel to another and from one instrument to another, the adjustment must be repeated.
Procedure
- Connect the probe to the channel and set the channel attenuation to match the probe’s selector.
- Connect the tip to the oscilloscope’s calibration signal terminal and the ground clip to its reference.
- Adjust the vertical scale and the time base to see several periods of the square wave, with a comfortable amplitude on the screen.
- With a small insulated tool, turn the probe’s compensation adjustment until the top of the wave is flat and the corners are square.
- Repeat when moving the probe to another channel or oscilloscope.
What each case looks like
| Observed shape | Condition |
|---|---|
| Corner with a peak or overshoot and a top sloping downward | Overcompensated |
| Rounded edges and slow rise | Undercompensated |
| Clean edges and a flat top | Compensated |
What is affected if it is skipped
A poorly compensated probe distorts the amplitude of fast signals and rise times: a clean pulse appears with ringing or with slower edges. Amplitude measurements of low-frequency sine waves barely change, so the error goes unnoticed until edges or digital signals are measured.
Common mistakes
- Compensating with a long ground clip: the ground lead adds inductance, and the result on the screen looks different from the measurement with a short ground spring.
- Compensating and then changing channels without repeating.
- Trying to compensate a probe in 1x mode, which has no such adjustment.
- Using a metal screwdriver, which disturbs the adjustment and the reading.
- Forgetting that compensation is no substitute for bandwidth: a probe cannot show what its bandwidth does not allow.
Caution
Compensation is done with the oscilloscope’s own low-level signal. Measuring mains or power circuits requires probes with the appropriate measurement category and qualified personnel who follow the applicable safety regulations.