What it covers in general terms
The standard deals with the safety of the equipment and of the person using it. It covers measuring instruments, control equipment and electrically powered laboratory apparatus. It is a generic standard: other parts of the series add requirements for particular types of equipment.
What hazards it considers
- Electric shock from accessible live parts.
- Mechanical hazards, such as edges, moving parts or instability.
- Excessive temperatures and fire spread.
- Radiation, liquids, pressure and other hazards specific to the equipment.
- Hazards arising from single faults, abnormal conditions or reasonably foreseeable misuse.
How it approaches safety design
The approach is that the equipment must remain safe after a single fault. It relies on concepts such as basic insulation, reinforced insulation, clearance and creepage distances, and protective earthing. These distances depend on the working voltage, the expected transient overvoltages and the pollution degree of the environment.
Measurement categories
The standard defines categories according to the point in the installation where an instrument is connected, from signal circuits up to the service entrance. This determines what transient overvoltage it must withstand. A low-category instrument is not suitable for a point in the installation with more energy available, even if its rated voltages seem sufficient.
What the engineer sees in practice
It shows up in the instrument markings: category symbols, maximum voltage to ground, insulation class and warnings. It also shows up in test leads with shrouds, finger guards and fuses with adequate breaking capacity. Choosing accessories that do not match the instrument’s category reduces the safety of the whole setup.
When it does not apply
It does not replace installation regulations or work procedures. For exact requirements, limits and test methods, consult the current edition of the standard. Work on energized circuits is the responsibility of qualified personnel in accordance with the applicable regulations.