A source and a meter in the same channel
An SMU forces one quantity, voltage or current, and measures the other. If it forces voltage, it measures the resulting current, and vice versa. Because both functions share the same channel and the same connection point, the errors of using two instruments with different cables and references are eliminated.
The compliance limit
When forcing voltage, the SMU limits the current to a compliance value; when forcing current, it limits the voltage. This protects the device under test, for example a diode or a transistor, and is essential for sweeps. If the instrument reaches the limit, the forced quantity is no longer the programmed value. This condition is usually flagged and must be checked when interpreting data.
Four-quadrant operation
Unlike a conventional power supply, which only delivers energy, an SMU can also absorb it. It works in the positive and negative quadrants of voltage and current, so it can both source and sink power. This allows it to trace the complete curve of a device, even in the region where the device acts as a generator, such as a photovoltaic cell.
Four-wire measurement
At high currents, the drop in the cables distorts the voltage reaching the device. An SMU usually has separate force and sense terminals: current flows through the former, and the voltage is measured directly at the device through the latter, with practically zero current.
What you see in practice
- Current-voltage curves of diodes, transistors, sensors and nonlinear resistors.
- Measurements of very small currents, where cable leakage, noise and humidity dominate.
- Sweeps with programmed delays to wait for the device to stabilize thermally.
How to interpret the result
A fast sweep can heat the device and distort the curve; a slow one reduces the effect but lengthens the test. Always compare the curve obtained against the compliance and integration time settings chosen. When hazardous levels are involved, connections are the responsibility of qualified personnel in accordance with the applicable safety regulations.