What is a data acquisition system?

A data acquisition system conditions, samples and digitizes signals from sensors or instruments, then stores them or sends them to a computer for logging, analysis or control.

What it is for

A data acquisition system converts physical quantities such as temperature, voltage, strain or pressure into digital data traceable over time. There are several stages between the sensor and the data, and errors accumulate at each one.

Stages in the chain

  1. Sensor or transducer: generates a voltage, current, resistance or charge proportional to the quantity.
  2. Conditioning: amplifies, filters, isolates or provides excitation to the sensor. A thermocouple needs cold-junction compensation and a strain gauge bridge needs excitation.
  3. Multiplexing: many channels share a single converter, which reads them sequentially. In other designs each channel has its own converter and all channels are sampled simultaneously.
  4. Analog-to-digital conversion: the resolution in bits determines the smallest distinguishable step within the selected range.
  5. Transfer and storage: over an equipment bus or a network, with a timestamp.

What you see in practice

If channels are read in sequence, there is a time offset between one and the next; on fast signals this shows up as a phase error between channels. Mains noise and multiple grounds appear as a periodic component on the signal. A reading that jumps when a cable is moved usually points to a shielding or connection problem.

What to define before measuring

  • Input range: if it is too wide, resolution is lost; if it is too narrow, the input saturates.
  • Sampling rate: it must be sufficient for the frequency content of the signal, and an anti-aliasing filter ahead of the converter prevents fast components from appearing as false slow signals.
  • Isolation: galvanically separates the signal from the electronics and prevents ground loops.
  • Total uncertainty: adds together the contributions of the sensor, the conditioning and the converter, not just the converter alone.

When it does not apply

For very high-speed single-shot events, an oscilloscope is usually more suitable, and for a single precise laboratory measurement, a high-accuracy multimeter. With hazardous voltages or mains signals, the connection must be made by qualified personnel following the applicable regulations.