The idea behind the measurement
The time interval between two events is the difference between the instant one occurs and the instant of the other. A counter with a time interval mode uses one channel to start the measurement and another to stop it; between the two instants it counts pulses from its time base, or uses interpolation to refine the resolution. An oscilloscope displays both signals and lets you measure the separation between the chosen points with cursors or an automatic measurement.
Steps
- Assign the channels: the initial event to the start channel and the final event to the stop channel.
- Define the edge of each event: rising or falling.
- Set the trigger level on each channel, preferably at a point on the signal where the slope is steepest, because there noise produces less uncertainty in the instant. A common criterion is half the amplitude.
- Choose an impedance and coupling for each input that are compatible with the signals.
- Average several measurements if the interval repeats, to reduce noise.
Sources of error
- Unequal cable delays: if the two cables or the two probes have different lengths or delays, the difference is added to the result. This is compensated by using identical cables or by measuring the delay with a common signal applied to both channels.
- Different trigger levels on signals with slow edges: the instant depends on the level chosen.
- Noise and jitter on the signals, which vary the trigger instant from one measurement to the next.
- Instrument resolution and the stability of its time base.
- Loading of the signals by the instrument input, which can change the transition time.
When to use each instrument
A counter offers higher resolution and statistics over many measurements; an oscilloscope lets you see the shape of the signals and detect problems such as edges with bounce. If the interval is very short or the edges are complex, the oscilloscope helps you decide where to measure.
How to interpret the result
A large spread between measurements indicates noise or jitter; a constant shift indicates cable delays or trigger levels. Swapping the channels makes it possible to separate the instrument’s bias from the real one.