How a counter counts
A frequency counter counts how many cycles of the signal occur during a known interval, called the gate time, and divides the number by that time. The interval is defined by an internal time base, which is an oscillator. That is why the accuracy of the measurement cannot be better than that of the oscillator, and the resolution depends on how long the gate lasts: a longer gate counts more cycles and distinguishes smaller differences, in exchange for a slower result.
Procedure
- Connect the signal to an input with an appropriate cable. For high-frequency signals, use coaxial cable and the input impedance that matches the source.
- Set the coupling and the attenuator so that the signal is within the input range without saturating.
- Set the trigger level at a point where the signal crosses cleanly, avoiding noisy zones. With noise on the signal, hysteresis or the input filter prevents false counts.
- Choose the gate time according to the required resolution.
- Consider an external reference if greater stability than that of the internal oscillator is needed; counters usually have an input for this.
Reciprocal counters
Modern counters measure the period of the signal with a fast time base and calculate the frequency as its reciprocal, which gives good resolution at low frequencies as well, where direct counting would yield few cycles per gate.
Sources of error
- Drift and aging of the reference oscillator.
- False triggering from noise or a poorly set level.
- Loading of the source by the input impedance.
- Signals with harmonics or modulation, which can make the counter count more crossings than expected.
How to interpret the result
If the reading fluctuates, increase the gate time or check the triggering. If it is stable but different from what was expected, compare with a known reference to evaluate the time base. A result is only as good as the reference behind it.