Definition
The components of an instrument, such as resistors, voltage references, amplifiers and oscillators, change their properties slightly with temperature. As a result, the instrument’s zero or gain shifts slowly and continuously as the internal or ambient temperature changes. This shift is called thermal drift.
How it is specified
It is expressed as a temperature coefficient: how much the error changes per degree of variation, in units such as parts per million per degree Celsius, or percent per degree. It is usually given separately for offset, which shifts all readings equally, and for gain, which scales them proportionally. An instrument’s accuracy specifications apply within a temperature interval around the calibration temperature; outside it, the temperature coefficient term must be added.
Where it comes from
- Voltage reference: this is the determining component in a multimeter or converter, and its thermal stability sets much of the performance.
- Internal resistors: changes in them affect attenuators and dividers.
- Self-heating: at power-on, dissipation raises the internal temperature until it reaches equilibrium.
- Environment: air currents, sunlight, proximity to equipment that gives off heat.
- Thermal electromotive forces: junctions of dissimilar metals in the connections generate voltages that depend on temperature gradients.
What the engineer sees
A reading of a stable reference that slowly rises or falls during the first hour after the instrument is switched on, or that changes when a door is opened or the air conditioning is turned on. In low-level measurements, the zero moves with the heat of a hand touching a connection.
How to limit it
- Let the instrument warm up for the time recommended by the manufacturer.
- Keep the ambient temperature stable and close to the calibration temperature.
- Zero the instrument or run a self-calibration when it offers one, and repeat it if the temperature changes.
- Protect connections from air currents and gradients.
- Record the temperature along with the readings so that corrections can be made later.