What is the input impedance of an instrument?

The input impedance of an instrument is the load it presents to the circuit being measured; if it is not much higher than the source impedance, the instrument itself alters the signal it is meant to measure.

Definition

Every measuring instrument takes some energy from the circuit it is connected to. Input impedance describes how much opposition it offers to that connection. It is modeled as a resistance in parallel with a capacitance: the resistance dominates at DC and low frequency, and the capacitance gains importance as frequency rises, because its reactance decreases.

The loading effect

The source being measured has its own output impedance. Together with the instrument’s input impedance, it forms a voltage divider. The instrument sees only a fraction of the real signal: the one given by the ratio of its input impedance to the sum of the two. If the input is much higher than the source impedance, the fraction is almost one and the error is negligible. If they are comparable, the error is large.

What it means in practice

  • Voltmeters and multimeters: they usually have a high input impedance to minimize loading. Even so, measuring a high-impedance source, such as a resistive divider with large values or a sensor with a weak output, can produce low readings.
  • Oscilloscopes: many offer a high-impedance input with a small capacitance and another of 50 ohm for sources and cables of that impedance. Choosing the wrong input distorts the waveform.
  • Test probes: an attenuating probe increases the resistance and reduces the capacitance seen by the circuit, at the cost of attenuating the signal.
  • Ammeters: they work the other way around; a low input impedance is wanted so as not to alter the current.

How it is detected

If the reading changes when the instrument is connected, or if a fast signal appears with softened edges, there is a loading effect. A simple check is to measure with another instrument of different impedance, or to add a known series resistance and observe the change.

How to interpret the result

Before measuring, it is advisable to know the source impedance and the instrument’s input impedance and to calculate the divider error. At high frequency, also compare the input capacitance with the node capacitance. Input impedance is a property of the instrument, but its effect on the measurement always depends on the circuit.